2003 Abstracts
A CLINICAL COMPARISON OF PORTABLE OXYGEN SYSTEMS: CONTINUOUS FLOW COMPRESSED GAS VS. OXYGEN CONCENTRATOR GAS DELIVERED WITH AN OXYGEN CONSERVING DEVICE
J. Lewarski BS RRT
1, G. Mikus RRT
1, G. Andrews BS RRT
1,
R. Chatburn BS RRT FAARC
2 1Hytech Homecare, Mentor OH;
2University
Hospitals of Cleveland, Cleveland OH
Background: Declining home oxygen reimbursement along with growing demand for small ambulatory oxygen (O
2) systems has encouraged the development of new ambulatory O
2 technologies. Recently, O
2 concentrators capable of filling small compressed gas cylinders have entered the market. Although it is well established in literature that O
2 produced from concentrators at >88% delivered in continuous flow is clinically equivalent to 99.6% USP gas (traditional compressed oxygen), there are no data evaluating concentrator gas delivered via an oxygen-conserving device (OCD). To test the hypotheses of equivalent benefit, we compared patient responses and tolerance of continuous flow (CF) USP O
2 versus compressed concentrator gas delivered via OCD in current home oxygen users.
Methods: We selected the HomefillÒ II oxygen concentrator and transfill system (Invacare, Elyria, Ohio), which includes a proprietary ML6 cylinder configured with a pneumatic OCD (EasyPulseÒ, Precision Medical, Northhampton PA). We used 9 patients in a prospective, randomized, crossover design. Patients were selected from a pool of existing home O
2 users from one home medical equipment provider. All patients were previously diagnosed with uncomplicated COPD and regularly using an ambulatory O
2 system. Additional selection criteria included: stable condition, O
2 prescription of 3 LPM or less and ability to carry portable. Physician orders were obtained for each patient. Patients were randomly assigned one of the following delivery systems: CF 99.6% O
2, or 93% O
2 concentrator gas via the ML6 with the OCD. Liter flow and settings for O
2 were consistent with their current prescription. On different days, each patient underwent 1 of the 2 test walks with the selected delivery system. A standard 6-min walk protocol was used while S
PO
2 and heart rate were continuously monitored and recorded. Objective measures of S
PO
2, HR, and distance walked along with subjective determination of breathlessness using a Borg Scale were used to evaluate the patient condition before and after each walk. Physiologic data were compared via 2-way ANOVA. The Borg Scale data was analyzed via Wilcoxon Rank Sum Test. A power analysis was performed for an effect size of 10% change in S
PO
2 and 15 beats/min for HR.
RESULTS: All patients tolerated the test. The table shows mean (standard deviation):
|
Device |
SpO2 |
Heart Rate |
Borg Score |
|
Before |
After |
Before |
After |
Before |
After |
|
99.6% O2 constant flow |
97% (1) |
90% (7) |
81 (13) |
112 (15) |
85 (1) |
111 (2) |
|
93% O2 with conserver |
96% (1) |
88% (11) |
83 (11) |
115 (10) |
85 (1) |
111 (1) |
There was no effect of device on either SpO
2 or heart rate (p=0.792). Statistical power was > 0.90. There was no difference in Borg score (p = 0.63).
Conclusions: These results suggest that the lower percentage O
2 output by the concentrator system does not adversely affect clinical outcomes when using an O
2-conserving device. O
2 derived from a concentrator at 93% O
2 and delivered in conjunction with a pneumatic O
2-conserving device provides the same clinical benefit as the standard 99.6% O
2 continuous flow device. Practical benefits of a transfilling oxygen concentrator system include patient freedom to refill their compressed gas cylinders at their own schedule, leading to improved portability. Providers should experience a substantial decrease in the high and recurring operational costs associated with the provision of ambulatory O
2 systems.