 |
PATIENT TRIGGER TO RESPONSE
TIME OF THE DRÄGER BABYLOG AND THE BIRD GOLD INFANT VENTILATORS IN A NEONATAL
BENCH STUDY
Jim Keenan BS, RRT, FAARC,
Primary Children?s Medical Center, Salt Lake City, Utah
Introduction: When ventilating
spontaneously breathing neonates it is extremely important for patient comfort
and synchrony for ventilator breaths to be delivered as quickly as possible.
Ventilator manufacturers refer to this as response time and many have different
ways of calculating and reporting it. For our purpose, we defined response
time as how quickly a ventilator begins to deliver flow at the endotracheal
tube after a patient?s spontaneous breath effort is detected by the ventilator.
We sought to measure response time of the Bird Gold and the Dräger Babylog.
Method: The Vitaltrends plethysmograph
isolette is a closed system that measures volumes and pressures displaced within
the system. The Vitaltrends uses a differential pressure transducer to measure
pressures and uses a single screen (300 mesh stainless steel) low resistance
pneumotachometer to measure volume. The plethysmograph?s pressure transducer
is wyed into the ventilator circuit?s proximal pressure line distal to the wye
piece. This allows graphic and waveform analysis on ventilated infants <3Kg.
The Vitaltrends computer package has the ability to measure and analyze graphic
waveforms in milliseconds. The Bird Gold and Dräger Babylog ventilators were
used with neonatal circuits ventilating an Ingmar neonatal test lung in separate
tests. New flow sensors were used for each experiment. The ventilator settings
used were: time cycled, pressure limited AC, Pressure 25 cm H2O, rate 20, flow
6 LPM, Ti 0.35 sec, and a trigger sensitivity of 0.3 LPM. A small plastic t-piece
was inserted between a 3.0 mm endotracheal tube?s connector and the tube itself.
The t-piece was connected to a 10 cc Hans Rudolph syringe. Fish line was attached
to the syringe plunger and was run out of the closed plethysmograph. The plethysmograph
was calibrated before each test as per the manufacturer?s specifications. A
spontaneous breath was simulated when the line attached to the syringe plunger
was pulled. Twenty spontaneous breaths were recorded and analyzed for both ventilators.
Results: To determine response
time we measured from the Vitaltrends graphics at the point of negative pressure
deflection, caused by our spontaneous trigger, to the first rise detected in
flow. Response time results are reported in the table.
Conclusion: Response time
is an important component for patient comfort and synchrony when choosing a
ventilator for neonates. The % difference indicates that the Dräger response
time is 0.28 shorter than Bird Gold, however, the actual time difference is
0.015 sec. Two questions that require further investigation are: 1) is a difference
of 0.015 sec. clinically significant between two devices? 2) What are appropriate
response times of ventilators for spontaneously breathing neonates?
| |
Bird Gold |
Dräger Babylog |
% Difference |
| Mean |
0.054 sec |
0.039 sec |
0.28 |
| Std Dev. |
0.016 |
0.008 |
|
P<0.008
Mann-Whitney U
OF-02-121
|
 |