 |
Bench Test of the Dräger Babylog
8000plus Neonatal Ventilator with Heliox Mixtures.
Melissa K. Brown RRT,
Sharp Mary Birch Hospital for Women, San Diego, California.
Background: Helium-oxygen
(heliox) gas mixtures are of a lower density than air/oxygen mixtures. Inhalation
of gas with a low-density such as heliox can reduce airway resistance. This
enables improved ventilation and the potential to increase expiratory flow and
to decrease lung hyperinflation and auto-PEEP. There are several studies documenting
the beneficial effects for neonatal patients with heliox therapy. I sought to
determine if the Dräger Babylog 8000plus ventilator could deliver heliox via
the 50-psi air connection of the ventilator. Also, the feasibility of leaving
the flow sensor in line to potentially enable patient triggering and synchrony,
and the accuracy of the oxygen blender and analyzer in the presence of heliox.
In addition, what technical difficulties might be encountered, due to the degree
of inaccuracy in volumes measured by the hot wire anemometer flow sensor?
Methods: An 80/20-heliox
mixture was connected via the 50-psi air inlet of the ventilator. The SIMV and
A/C modes were utilized, pressures 20/5, r 30, 0.4 inspiratory time with heliox
mixtures in 5-10% increments from 10/90 to 79/21. FiO2 was measured
externally with a Maxtec MaxO2 electrochemical analyzer (Salt Lake
City, Utah). Volumes were measured with the Novametrix Co2smo plus
Respiratory Mechanics Monitor (Wallingford CT), which contains software programmed
for compensation of heliox mixtures. The hot wire flow sensor was left in line
for all concentrations except 75/25 and 79/21, when the high minute alarm maximal
threshold was breached (15 L/min) and could not be disabled.
Results: The Dräger
Babylog 8000plus cycled consistently with all heliox mixtures. Once heliox concentrations
reached 50/50 the ventilator flow sensor was no longer able to display any tidal
volume on the test settings and the apnea alarm had to be disabled. When heliox
concentrations reached 75/25 the flow sensor had to be removed enabling IMV
ventilation only. The ventilator oxygen analyzer was within 1% of the external
analyzer, but the blender was as much as 9% off causing audio and visual alarms,
due to internal alarm limits of ± 4% of the internal oxygen analyzer. The ventilator
flow sensor tidal volumes were either erroneous or unable to read any volumes,
at all concentrations of heliox.
Conclusion: The Dräger
Babylog 8000plus gas mixing system has a zero internal bleed flow, enabling
heliox mixtures to be introduced through the 50 psi air inlet, without wasting
gas and allowing consistent cycling of the ventilator. The ventilator oxygen
analyzer is accurate, but has to be disabled to overcome the constant alarm
condition caused by the disagreement between blender setting and FiO2 delivered. The flow sensor can be left active to facilitate patient triggering
and synchrony until the ventilator flow sensor measures exhaled minute volumes
of 15 L/min. Tidal volume measurements by the ventilator flow sensor are erroneous
or absent and a supplemental monitor such as the Co2smo plus should
be utilized. Once the flow sensor is removed, IMV ventilation with heliox can
be provided. More research should be done evaluating the ability of the ventilator?s
flow sensor to accurately detect neonatal triggering efforts with heliox gas
mixtures.
OF-02-049
|
 |