 |
An
Investigation to Determine the Compatibility of the Hamilton Galileo Ventilator
and the Jet Bunnell in a Neonatal Lung Model
Jim Keenan BS RRT,
Troy Lynch RRT. Respiratory Care Service, Primary Children?s Medical Center,
Salt Lake City, Utah.
Introduction: Our facility
is in the process of reevaluating the standard of care ventilator for our level
III NICU. The Bunnell Jet ventilator is often used in tandem with our current
standard ventilator for patients meeting jet ventilation criteria. The reevaluation
process must insure that whatever ventilator standard is chosen, that it is
compatible with the Jet Bunnell. Hamilton Medical Inc. has recently come out
with neonatal software upgrades for their Galileo ventilator. The package includes
time-cycled pressure-limited modes with a neonatal flow sensor. Since this ventilator
is one of many that is being considered we decided to do a bench test to determine
its compatibility with the Jet Bunnell since it had not been previously reported
in the literature.
Methods: Testing consisted
of ventilating an IngMar Medical Neonatal test lung using the Galileo and the
Jet in tandem under two test conditions. Each test condition was performed
using a triple lumen 3.0 and 4.0 ETT. In test condition one, PIP was varied
on the Galileo from 5 to 35 cmH20 in 5 cmH2O increments. The jet PIP was operated
at 5 cmH2O higher than the Galileo and also increased in 5 cmH2O increments,
along with the Galileo, to 40 cmH2O. All other ventilator settings remained
the same. These setting were: Galileo: time cycled - pressure limited SIMV,
rate: 5, Peep; 5, inspiratory time: 0.4 sec., flow: 6 Lpm, trigger 1.5 Lpm.
Jet: rate 420 with an on-time of 0.02 sec. In test condition two the Jet rate
was varied at rates of 300, 420, and 500. All other ventilator settings remained
the same. These setting were: Galileo: time cycled - pressure limited SIMV,
PIP: 20 cmH2O, rate: 5, Peep; 5, inspiratory time: 0.4 sec., flow: 6 Lpm, trigger
1.5 Lpm. Jet: PIP: 25 cm H2O with an on-time of 0.02 sec. Data was gathered
for approximately 5 min. for each variable test condition. All measured and
monitored Galileo data was captured using the Galileo Data Logger software package.
Results: Both ventilators
operated efficiently under these test conditions. All measured and monitored
data from the Galileo were consistent, reproducible, and within acceptable limits.
Upon initial investigation under test condition one, it was discovered that
the PIP of the Galileo must always be at least one cmH2O below the PIP of the
Jet. If not, the Galileo would report a sensor malfunction. During the alarm
condition the Galileo continued to function appropriately. This may be of concern,
as some clinicians prefer to ventilate with conventional PIP?s the same or higher
than the jet.
Discussion: When new neonatal
ventilator technology becomes available it is important to consider all aspects
of its operation. Those facilities that promote the use of Bunnell Jet ventilation
must consider if the two devices are compatible. In addition, many NICU?s are
recognizing the benefit of flow sensor monitoring technology. The sensor and
often its location create more difficulties when the jet us used.
Conclusion: We feel that it
is safe to use the Hamilton Galileo in tandem with the Bunnell jet as long as
it is clinically acceptable to use jet PIP?s greater than the Galileo.
OF-01-231
|
 |