Homepage of Cardinal Health
Investor Relations     Careers     Our Businesses 
About Us  News   
Cardinal Health Medical Products and Services

 




Comparison of Bird VIP Gold Ventilator in Conventional Volume Mode to Dräger Babylog 8000+ In Volume Guarantee Mode, A Bench Study



E.Collins, RRT; V. Faciane, BS RRT; L Abbruzzese, RRT; Providence Hospital, Anchorage, Alaska

Background: The role of ?Volutrauma? in neonatal lung injury has received increased interest. Two ventilators studied offer different approaches to a volume-targeted breath. VIP Gold offers conventional volume ventilation with a decelerating flow. Babylog 8000+ offers pressure limited breaths (TCPL) (PSV) that adjust the pressure target to accomplish the selected tidal volume when ?Volume Guarantee? is active. Understanding how these modes function in various clinical settings is vital to their appropriate use.

Objective: To investigate the volume delivery of two ventilators to changes in Compliance and Resistance.

Design/

Methods:
Using an IngMar neonatal test lung, both ventilators were set to achieve a desired expiratory tidal volume at a compliance of 1.0 ml/cm. Compliance was changed to .5, 1.5 2.0; returned to 1.0 ml/cm and the resulting tidal volume and PIP recorded for 30 breaths. The process was repeated with a 2.5mm Et tube and a 3.5mm Et tube. The Babylog was set in Assist/Control TCPL with Volume Guarantee. The VIP Gold was set in Assist/Control volume mode with decelerating flow. Both ventilators used the same standard disposable heated wire circuit (Allegiance) and Fisher/Paykel 750 heated humidifier. Target tidal volumes were 5, 10, and 15 ml.

Results: Both ventilators delivered variable expiratory tidal volumes when compliance was changed. (See figure below) The Dräger Babylog 8000+ delivered large fluctuations in expiratory tidal volumes. The VIP Gold had less fluctuation in tidal volume but did not return entirely to the set tidal volume. Maximum overshoot: Dräger Babylog 150%; VIP Gold 37%. Changes in resistance did not change the expiratory tidal volume significantly for either ventilator.

Conclusion: Changes in compliance resulted in significant changes to actual measured expiratory tidal volume. Such sudden compliance changes could occur in response to surfactant therapy, when an Et tube is repositioned or a pnuemothorax is recognized. Volume overshoot with the Dräger Babylog may be attributed to the TCPL mode. A compliance change could result in a larger volume delivery before the ventilator re-adjusts the pressure target. VIP Gold volume changes may be attributed to circuit compliance change. Both ventilators monitor delivered tidal volume at the wye. The results of the bench test warrant further study.



OF-01-202
   Contact Us     Legal/Privacy Policy     Suppliers & Distributors     OEM Services     Search     Home